Automatic 100kv Transformer Oil Breakdown Voltage Test Machine IEC156

Model NO.
GDYJ-502
Display
Digital
Control
Fully Automatic
Weight
41kg
Power Source
AC220V
Maximum Storage Capacity
100 Groups of Test Data
Test Times Setting
1-9
Working Environment
0-50ºC,≤80%Rh
Power Supply
AC220V±10%, 50Hz/60Hz
Rate of Voltage Rise
2kv/S±5%, 3kv/S±5%,5kv/S±5%
Measuring Precision
3%Rdg+0.3%Fs
Output Voltage
AC 0~80kv
Standards
IEC60156, Is6792, BS5874, ASTM D1816, ASTM D877
Transport Package
Wooden Case
Specification
620*430*330mm
Trademark
GOLD
Origin
Chongqing
HS Code
9031809090
Production Capacity
300PCS/Month
Reference Price
$ 1,215.00 - 1,350.00

Product Description

Automatic 100kv Transformer Oil Breakdown Voltage Test Machine IEC156

Automatic 100kv Transformer Oil Breakdown Voltage Test Machine IEC156
Introduction:
  • Automatic 100kv Transformer Oil Breakdown Voltage Tester is used for transformer oil / insulating oil dielectric strength / breakdown voltage testing  
  • It has large LCD display, build-in micro-printer and 100 groups data can be stored. Also   you can upload the test data to PC from RS232 for data analyzing.

  • The special circuit design makes the tester has a good performance at stability and anti-noise.

Functions:
  • Breakdown test of user-defined;

  • Proof test;

  • IEC60156/IS6792/BS5874 test;

  • ASTM D1816 standard test;

  • ASTM D877 standard test;

  • 5 minutes standard test;

  • Manual test;

  • Calibration interface for customers;

  • Uploading data to PC and making word report automatically;

  • Result print;

Specifications:

1 Output voltage AC 0~80KV/AC 0~100KV
2 Rate of voltage rise 0.5kv/s ± 5%, 2kv/s ± 5%, 3kv/s ± 5%,5kv/s ± 5%
3 Measuring precision 3%RDG+0.3%FS
4 Test times setting  1~9
5 Mixing time setting  0~99 second
6 Waiting time setting 0-9 minute 59second
7 Maximum storage capacity 100 groups of test data
8 Test standards IEC60156, IS6792, BS5874, ASTM D1816, ASTM D877
9 Display 256*128 LCD
10 Rated power supply AC 220V±10% 50Hz/60Hz
11 Environment temperatures  0-50°C   
12 Relative humidity ≤80%RH
13 Size    620 mm*430mm*330mm
14 Weight 41kg
 

GDYJ Series   Transformer Oil Breakdown Voltage Tester

Model

GDYJ-501

GDYJ-502

GDYJ-502A

Output Voltage

AC 0~80KV/0~100KV

AC 0~100KV

AC 0~80KV/0~100KV

Rate of Voltage Rise

2KV/s±10%

0.5/2/3/5 KV/s±5%

0.5/2/3/5 KV/s±5%

Max. Time Setting

0~9mins

9mins

0~99 seconds

Test Precision

3% RDG+0.3%FS

3% RDG+0.3%FS

1% of full scale +2digits

Max. Data Storage

100 groups

100 groups

100 groups

Power Supply

AC220V±10% 50/60Hz

AC220V±10% 50/60Hz

AC220V±10% 50/60Hz

Environment Temp.

0~50Centigrade

0~50Centigrade

0~50Centigrade

Related Humidity

<80%RH

<80%RH

<80%R H

Dimension

420*380*410mm

620*430*330mm

490*410*450mm

Weight

28kg

41kg

45kg


Product Pictures:

Automatic 100kv Transformer Oil Breakdown Voltage Test Machine IEC156 Automatic 100kv Transformer Oil Breakdown Voltage Test Machine IEC156
Testing Process:
Automatic 100kv Transformer Oil Breakdown Voltage Test Machine IEC156

 

PNEUTEC.IT, 2023