Parameters | |
Display mode
|
HD 10.4 TFT color LCD display
|
Ventilation mode
|
IMV, VCV, PCV, PSV, CPAP,SIMV,PSV+SIMV,
|
Ventilation function | PEEP,SIGN,standby,IP,IRV |
ventilation parameters |
Tidal volume:
20-1800 ml
Rate: 2-120bpm SIMV rate: 2-20bpm I:E: 4:1-1:8 Pressure trigger sensitivity :-10-10cmH2O Flow trigger sensitivity :1-20L/min PEEP: 0-25cmH2O Pressure range: 3-50cmH2O Pressure limitation :5-60cmH2O Pressure slope: 1-10gear Flow rate setting: 1-10gear Expiratiry trigger sensitivity :10%-90% Time control: 1-10s SIGH :1.5times the inspiratory time(60-120 adjustable) Inspiratory platform :0-50%
Oxygen concentration
of 21% to 100%
|
Ventilation monitoring | Expired tidal volume, Inhale tidal volume, Minute Ventilation volume, IPPV rate,SIMV rate, total respiratory rate, autonomous breathing rate, I/E, peak pressure of airway, average pressure, PEEP, Inspiratory trigger pressure,inspiratory platform, oxygen concentration, battery capacity , pressure - time waveform, flow rate - time waveform, flow-volume loop, pressure -volume loop |
Security Alarm System |
*Oxygen concentration alarm
upper limit setting range:21-99%
low limit setting range:18%-80%
*Airway pressure alarm
upper limit setting range:5-60cmH2O
Low limit setting range:0-50cmH2O
*Minute ventilation volume alarm
upper limit setting range:1-20L/min Low limit setting range:0-19L/min *Tidal volume upper limit setting range:10-1800ml Low limit setting range:0-1700ml *Respiratory rate upper limit setting range:10-99bpm Low limit setting range:0-50bpm *Sustained high-pressure alrm it will give alarm when stress have consistently been higher than 30 cmH2O
*Suffocation alarm
:power alarm
*Low pressure of Oxygen alarm :intubation off alarm *Low pressure of Air alarm :battery voltage alarm |
Power Supply |
AC 220V ± 10% 50Hz ± 1Hz
UPS power supply, Storage battery
|
Air compressor | With 'standby"function |
Mechanical Arms |
THR-AV-2000B1 THR-AV-2000B2 THR-AV-2000B3