Ventilation modes:
|
IPPV; A/C; PCV; SIMV; SIGH ; MAUN |
| Ventilator parameter ranges |
|
| Flow meter | O 2 (0.1-10L/min) |
|
|
N 2 O(0.1-10L/min) |
|
|
AIR(0.1-10L/min) |
| Rapid oxygen supply | 35L/min-75L/min |
| Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
| Frequence ( Freq ) | 1bpm ~ 100 bpm |
| I/E | 4:1 ~ 1:8 |
| PEEP | 0cmH2O ~ 30 cmH2O |
| Pressure triggering sensitivity ( PTr ) | - 20 cmH2O ~ 0 cmH2O ( Based on PEEP ) |
| Flow trigger sensitivity ( FTr ) | 0.5 L/min ~ 30 L/min |
| Pressure control (PC) | 5 cmH2O ~ 60 cmH2O |
| SIGH | 0(off) 1/100 ~ 5/100 |
| Apnea Ventilation | OFF , 5 s ~ 60 s |
|
Pressure Limit
|
20 cmH2O ~ 100 cmH2O |
| Monitored Parameters |
|
| Frequence ( Freq ) | 0 /min ~ 100 /min |
| Tidal volume(Vt) | 0 mL ~ 2000 mL |
| MV | 0 L/min ~ 100 L/min |
| Oxygen concentration |
15 %
~
100 %
|
| Graphical display: |
|
| P-T(pressure - time) |
|
| F-T(flow - time) |
|
| P-V loop (pressure - volume loop) |
|
|
Size Wooden case packing size : L 770*W 890*H 1535mm , GW : 151KG ; NW : 91.2KG Alarm and protection |
|
| The AC power failure alarm | Power failure or no connection |
| Internal battery backup low voltage alarm | < 11.3±0.3V |
| No tidal volume | ≤5Ml within 6s |
|
High oxygen concentration alarm
Low oxygen concentration alarm |
19%-100%
18%-99% |
|
High Airway pressure alarm
Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm |
20cmH2O-100cmH2O
0cmH2O-20cmH2O Adult(5L/min-20L/min) Paed(1L/min-15L/min 0-10L/min) (PEEP+1.5kPa) over 16s |
| Suffocation warning | 5s-60s no spontaneous ventilation |
| The maximum limited pressure | < 12.5 kPa |
|
Fan error
Oxygen deficit |
Show on screen
Show on screen |
| Working conditions |
|
| Gas source | O2,N2O,AIR |
| Pressure | 280kPa-600kPa |
| Voltage | 100-240V |
| Power frequency | 50/60Hz |
|
Input power
|
85VA |
| Vaporizer |
|
| Anesthesia gas | Adjustable scope % (volume percentage) |
| Halothane | 0 ~ 5 |
| Enflurane | 0 ~ 5 |
| Isoflurane | 0 ~ 5 |
| Sevoflurane | 0 ~ 8 |