Ventilation modes
|
IPPV; A/C; PCV; SIMV; SIGH ; MAUN |
Ventilator parameter ranges | |
Flow meter | O 2 (0.1-10L/min) |
N 2 O(0.1-10L/min) | |
AIR(0.1-10L/min) | |
Rapid oxygen supply | 35L/min-75L/min |
Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
Frequence ( Freq ) | 1bpm ~ 100 bpm |
I/E | 4:1 ~ 1:8 |
PEEP | 0cmH2O ~ 30 cmH2O |
Pressure triggering sensitivity ( PTr ) | - 20 cmH2O ~ 0 cmH2O ( Based on PEEP ) |
Flow trigger sensitivity ( FTr ) | 0.5 L/min ~ 30 L/min |
Pressure control (PC) | 5 cmH2O ~ 60 cmH2O |
SIGH | 0(off) 1/100 ~ 5/100 |
Apnea Ventilation | OFF , 5 s ~ 60 s |
Pressure Limit
|
20 cmH2O ~ 100 cmH2O |
Monitored Parameters | |
Frequence ( Freq ) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2000 mL |
MV | 0 L/min ~ 100 L/min |
Oxygen concentration |
15 %
~
100 %
|
Oscillogr aphs display | |
P-T(pressure - time) | |
F-T(flow - time) | |
P-V loop (pressure - volume loop) | |
V-T(volume - time ) | |
Size Wooden case packing size : L 870*W 890*H 1600mm , GW : 190KG ; Anesthesia Machine size : L930*W750*H 1405 mm ,NW : 124KG Alarm and protection |
|
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | < 11.3±0.3V |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm
Low oxygen concentration alarm |
19%-100%
18%-99% |
High Airway pressure alarm
Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm |
20cmH2O-100cmH2O
0cmH2O-20cmH2O Adult(5L/min-20L/min) Paed(1L/min-15L/min 0-10L/min) (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | < 12.5 kPa |
Fan error
Oxygen deficit |
Show on screen
Show on screen |
Working conditions | |
Gas source | O2,N2O,AIR |
Pressure | 280kPa-600kPa |
Voltage | 100-240V |
Power frequency | 50/60Hz |
Input power
|
80VA |
Vaporizer |
Anesthesia gas | Adjustable scope % (volume percentage) |
Halothane | 0 ~ 5 |
Enflurane | 0 ~ 5 |
Isoflurane | 0 ~ 5 |
Sevoflurane | 0 ~ 8 |