Class II Biological Safety Cabinet BSC
Model:MC-BSC2
SPECIFICATION
| MODEL | ABS-1100IIB2 | ABS-1300IIB2 | ABS-1500IIB2 | ABS-V | |
|
External Size
(W*D*H) (mm) |
1100*750*2250 | 1300*750*2250 | 1500*760*2250 | 1820*750*2250 | |
|
Internal Size
(W*D*H) (mm) |
940*600*660 | 1150*600*660 | 1350*600*660 | 1670*600*660 | |
| Tested Opening | Safety Height ≤ 200 mm (8") | ||||
| Max Opening | 420 mm (17") | 420 mm (17") | 500mm (20") | 500mm (20") | |
| Inflow Velocity | 0.53±0.025 m/s | ||||
| Down flow Velocity | 0.33±0.025 m/s | ||||
| Pre-Filter | Washable | ||||
| HEPA Filter | Two, 99.999% efficient at 0.3 um. Filter life indicator. | ||||
| Front Window | Motorized. Two-layer laminated toughened glass ≥ 5mm. Anti UV | ||||
| Noise | NSF49 ≤ 61dB / EN12469 ≤ 58 dB | ||||
| Illumination | ≥ 1000Lux | ||||
| Fluorescent Lamp | 21W× 2 | 21W× 2 | 28W× 2 | 40W*2 | |
| UV Lamp | 30W*1 | 18W*1 | 40W*1 | 40W*2 | |
| UV timer, UV life indicator, emission of 253.7 nanometers for most efficient decontamination. | |||||
| Consumption | 700W | 850W | 900W | 1200W | |
| Waterproof Sockets | Two, total load of two sockets: 500W | ||||
| Display | LCD display: exhaust filter and downflow filter pressure, filter and UV lamp working time, inflow and down flow velocity, filter life, humidity and temperature, system working time etc. | ||||
| Control System | Microprocessor | ||||
| Airflow System | 0% air re-circulation, 100% air exhaust | ||||
| Visual and Audio Alarm | Filter replacement, window over height, abnormal air flow velocity. | ||||
| Exhaust Duct | 4 meters PVC duct, Diameter: 300mm | ||||
| Material | Work Zone: 304 stainless steel | ||||
| Main Body: Cold-rolled steel with anti-bacteria powder coating. | |||||
| Work Surface Height | 750mm | ||||