High-Capacity Hba1c Analyzer Rapid Test Hematolobin
LD-560
Fully Automated HbA1c Analyzer
provides comprehensive, fully automated HbA1c and HbA2/F testing with a compact footprint within 90 seconds per test. Designed for clinics, physician offices, and clinical laboratories with small to medium sample volumes of 10-50 daily samples, fits in almost any labs and saves valuable bench space.
Specification
Model
|
LD-560
|
Methodology
|
High-Performance Liquid Chromatography (HPLC)
|
Test Modes
|
Fast Mode, Variant Mode
|
Test Range
|
3% - 18%
|
Precision (CV)
|
<2%
|
Parameter
|
HbA1c, HbF, HbA1a, HbA1b, HbA0, HbA2(Optional)
|
First Sample Result
|
4 Mins
|
Test Speed
|
Fast Mode: 90 Seconds/Test
Variant Mode: 140 Seconds/Test
|
Sample Type
|
Whole blood, Pre-diluted blood
|
Sample Volume
|
Whole blood: 5μL
Pre-diluted blood: 150μL
|
Sample loading capacity
|
20
|
Photometer
|
420nm
|
Column Capacity
|
3000T
|
Filter Capacity
|
600T
|
Display
|
10" LCD touch screen
|
Reagent Kit
|
Eluent A, Eluent B, Eluent C, Hemolysis, Chromatography Column
|
Storage
|
5000 Sample Results
|
Connection interface
|
USB, LAN, LIS Compatible
|
Printer
|
Thermal Printer
|
System operation
environment
|
Temperature 15-30ºC
Humidity 30%-85%
|
Power
|
AC 100-240V 50/60HZ 300VA
|
Dimensions
|
378mmx380mmx510mm
|
Weight
|
28Kgs
|
Performance
HPLC Technology - Gold Standard Methodology Fully Automated - To Minimize Operation Hassles
• NGSP and IFCC certified • Primary Tube Sampling with Cap Piercing
• HbAlc results in 130 seconds • Fully Automated Start-up, Maintenance
• Barcode Scanner for Sample identification
Precise and Reliable - To Serve You Consistently
Compact Size - To Minimize Space Requirement
s
• HbAlc Inter & Intra Measuring CV's≤3%to • Small Footprint Reduces Bench Space Needed
EnableExceptional ResultManagement
• SuperiorQuality Chromatographic Resolution to
Eliminatelnterferences
HPLC Methodology
High-Performance Liquid Chromatography (HPLC), to separate HbA1c directly with measuring the absorbance points continually to form chromatogram. Using normal distribution curve fitting auto-iterative algorithm to get precise HbAlc testing result, excluding interference of variant and unstable hemoglobin like HbF. Standard Analysis Mode will report HbA1a, HbA1b, HbF, HbA1c, P3, HbA0 peak areas and ratio. And the result also includes IFCC, NGSP and eAG value for diverse client needs.
Reagents and Consumables
The analyzer is recommended to be used in conjunction with the Glycosylated Hemoglobin Test Reagent Kit (high performance liquid chromatography (HPLC method)) produced by our company, which includes the following components:
600T: Eluent A/Eluent B/
Eluent C(
Variant Mode
)
|
Adjusts the pH of the solution for the elution of glycosylated hemoglobin and non glycated hemoglobin.
|
600T:
Hemolysis Reagent
|
Used to dissolve red blood cells and release hemoglobin.
|
3000T:
Chromatography Column
|
Used for separation and elution of glycosylated hemoglobin and non glycated.
|
600T:
Hemoglobin Filter
|
Used to initially filter the sample after hemolysis.
|
12Pcs: Paper Rolls
|
Prints a report of the test results.
|
Certificate
Company Information
LABNOVATION TECHNOLOGOES, INC was established in 2001 and has been recognized as a "National High-tech Enterprise". The registered capital (including its wholly-owned subsidiaries) is RMB 20 million, and it is one of the earliest companies exporting in vitro diagnostic reagents in China. The product sales network has covered more than 110 countries and regions around the world.
It is the first in the industry to pass the ISO9001:2008.ISO13485:2012/AC:2012 quality system TUV certification, and obtain the EU CE certification and FDA certification.
The company invests more than 10% of its sales into product research and development, hardware construction and update every year, and 85% of the company's R&D team employees have more than 5 years of industry experience.