Double Stage RO System EDI Electric Desalination 500lph for Hemodialysis

Model NO.
KYRO-500LPH
Feed Water TDS
<2000mg/L
Output Conductivity
<0.2us/Cm
Desalination Rate
97%
Imported RO Membrane
USA Dow Lcle-4040
Recovery Rate
50% -75%
Multiple -Choice Control Valve
Automatic or Manual
Name
EDI Electric Desalination 500lph for Hemodialysis
Function
Water Purifying/ Drinking/ Inustrial
Transport Package
Standard Packing
Specification
500L/H
Trademark
Kaiyuan
Origin
China
HS Code
8421299090
Production Capacity
500L/H
Reference Price
$ 5,940.00 - 6,120.00

Product Description

Double Stage Ro System and EDI Electric Desalination 500LPH For Hemodialysis

Advantages of EDI Technology

1.Best stability of water quality

2.Constantly produce ultrapure water as per user requirements

3.Modular production, can achieve automatic control

4.No need acid and alkali regeneration, no pollution emissions

5.No need stop because of regeneration

6.No need regeneration equipment and chemical storage

8.Low operation and maintenance cost

9.Simple operation and low labor strength

EDI system

EDI, also known as electric desalination. It is a new ultra pure water technology which gradually developed in the 20 th  century. EDI equipment combined electro dialysis technology and ion exchange technology. It is using the selectivity penetrate function of anion and cation membrane and exchange function of ion exchange resin, under direct current electric field to make ion directed migration. Finally to make deep desalination of water. In the same time, hydrogen ion and hydroxide ion which produced when water ionize will regenerate ion exchange resin. So if use EDI equipment, no need acid and alkali chemical to do any regeneration, and it can produce ultra pure water continuously.

Application 

a. Medical production,industrial pharmacy use
b. Cosmetics,  Hemodialysis
c. Recycling water system
d. Chemical, electroplate industrial etc
e. Microelectronic, power, and biological project
f. Chemical Lab,semiconductor
g. Other electronic standard water

Process
 
Raw water tank(user supply) → raw water pump → sand filter → carbon filter → softener filter → security(micro) filter → high pressure pump 1 st  → reverse osmosis system 1 st  →high pressure pump-2 nd  reverse osmosis system 2 nd →UV lamp→EDI system→ pure water tank(user supply)   - outlet
 
Main parts function

1. Raw water pump-- provide the pressure to quartz sand filter/active carbon filter/softener.
2. Multi-medium filter--get rid of turbidity, suspended matter, organic matter, colloid, etc.
3. Active Carbon filter--Remove the color, smell, free chloride, organic matter, harmful matter, etc.
4. Softener--exchange "ions" of calcium and magnesium with the most common and easiest method.
5. Security filter--prevent large particles, bacteria, viruses into RO membrane, accuracy is 5um,
6. High pressure pump--Provide the high pressure to RO membrane.
7. RO system-- main part of the plant. The RO membrane's desalting rate can reach 99%, removing over 99% ions.

Double Stage RO System EDI Electric Desalination 500lph for Hemodialysis
Double Stage RO System EDI Electric Desalination 500lph for Hemodialysis
Double Stage RO System EDI Electric Desalination 500lph for Hemodialysis
Double Stage RO System EDI Electric Desalination 500lph for Hemodialysis
Packaging
Double Stage RO System EDI Electric Desalination 500lph for Hemodialysis
Double Stage RO System EDI Electric Desalination 500lph for Hemodialysis
Double Stage RO System EDI Electric Desalination 500lph for Hemodialysis


Electric control box

Double Stage RO System EDI Electric Desalination 500lph for Hemodialysis

 

Thank you so much for your value time.Any questions please click " contact now " send us inquiry !
     
 
Guidelines for getting a proper quotation
Tell us the  raw water /source of water
Provide water  analysis  report
Required  production capacity ( 500 liters per hour, 1000liters per hour…)
What's the  pure water used for ( drinking, bottle water or industrial use…)

 

PNEUTEC.IT, 2023